The sock electrode array: a tool for determining global epicardial activation during unstable arrhythmias
Pacing and Clinical Electrophysiology : PACE
|September 1, 1980
Summary
A novel sock electrode array enables rapid, simultaneous epicardial mapping of heart electrical activity. This technique overcomes limitations of conventional methods for studying transient arrhythmias.
Area of Science:
- Cardiovascular Electrophysiology
- Medical Device Engineering
Background:
- Conventional epicardial mapping is time-consuming (5-15 min for 50+ sites).
- This method is insufficient for transient or beat-to-beat varying arrhythmias.
Purpose of the Study:
- To develop and evaluate a "sock electrode array" for rapid, simultaneous epicardial mapping.
- To overcome limitations of traditional hand-held electrode techniques.
Main Methods:
- A contour-fitting sock with 26 or 52 electrodes was developed.
- Simultaneous recording of epicardial potentials and computer-generated isochronous maps.
- Comparison with hand-held electrode technique in six normal dogs.
Main Results:
- Sock electrode array allowed simultaneous recording from 26 electrodes in one cardiac cycle.
- Isochronous maps generated within 10 minutes, comparable to conventional methods.
- Hand-held technique required 25-45 minutes for data acquisition and analysis.
Conclusions:
- The sock electrode array provides rapid and accurate epicardial mapping.
- This method is suitable for intraoperative studies of ventricular tachyarrhythmias.
- Enables timely analysis of epicardial electrical phenomena during surgery.