Clinically significant effect of protein concentration on ion-selective electrode measurements of ionised calcium
Others, using an Orion SS-20 ionised calcium analyser, noted that the ionised calcium concentration of a native serum sample was 8% greater than that of its ultrafiltrate. The experiments described here, using a Nova 2 ionised calcium analyser, confirmed a positive protein interference which was greater for human albumin than for IgG. Uncharged dextran showed no positive interference but dextran sulphate, which is highly charged and binds calcium, showed a large effect. Thus the interference is related to macromolecular charge. Dialysis experiments with normal and pathological human serum samples indicated that the ionised calcium of diffusible plasma water was overestimated by an average of 9.6% at an albumin concentration of 40 g/l and by 4.8% at 20 g/l. It is concluded that the measurement of ionised calcium with existing analysers can be clinically misleading in patients with abnormal plasma proteins. Hypocalcaemia is likely to be overdiagnosed and hypercalcaemia underdiagnosed in the presence of hypoalbuminaemia.
Others, using an Orion SS-20 ionised calcium analyser, noted that the ionised calcium concentration of a native serum sample was 8% greater than that of its ultrafiltrate. The experiments described here, using a Nova 2 ionised calcium analyser, confirmed a positive protein interference which was greater for human albumin than for IgG. Uncharged dextran showed no positive interference but dextran sulphate, which is highly charged and binds calcium, showed a large effect. Thus the interference is related to macromolecular charge. Dialysis experiments with normal and pathological human serum samples indicated that the ionised calcium of diffusible plasma water was overestimated by an average of 9.6% at an albumin concentration of 40 g/l and by 4.8% at 20 g/l. It is concluded that the measurement of ionised calcium with existing analysers can be clinically misleading in patients with abnormal plasma proteins. Hypocalcaemia is likely to be overdiagnosed and hypercalcaemia underdiagnosed in the presence of hypoalbuminaemia.
Related Concept Videos
Common Ion Effect
Feedback Regulation of Calcium Concentration
Various transmembrane receptors, such as G protein-coupled receptors (GPCRs), elicit a response to extracellular signals by increasing cytosolic calcium. Activated GPCRs...
Ionic Strength: Effects on Chemical Equilibria
In this solution, the primary cation—the calcium...
Complexometric EDTA Titration Curves
Ion-Exchange Chromatography
Determination of Multiple Dosing Parameters: Steady-State, Minimum and Maximum Concentrations