Two filter-based methods accurately measure airborne ammonia, unaffected by humidity. Passive monitors showed increased uptake with higher air velocity, while a colorimetric badge system proved inaccurate and biased.
Area of Science:
- Environmental Chemistry
- Analytical Chemistry
- Occupational Health
Background:
- Accurate measurement of airborne ammonia is crucial for environmental monitoring and occupational safety.
- Existing methods require evaluation for reliability and practicality in diverse industrial settings.
Purpose of the Study:
- To evaluate and compare the accuracy and performance of three different methods for measuring airborne ammonia.
- To assess the influence of environmental factors like humidity and air velocity on measurement accuracy.
Main Methods:
- Two methods utilized filters impregnated with phosphoric acid, analyzed via ion chromatography or colorimetry (Nessler).
- Sampling was performed using active filter cassettes and passive monitors (Gasbadge).
- A third method, a colorimetric air-monitoring badge system (ProTek), was also evaluated.
Main Results:
- Filter-based methods demonstrated good accuracy and were not significantly affected by humidity.
- Ion chromatography analysis showed no interference from amines.
- Gasbadge passive monitors exhibited increased ammonia uptake at higher air velocities (0.2 to 1.0 m/s).
- The ProTek method showed poor accuracy, bias, and high blank values.
Conclusions:
- Impregnated filter methods, particularly when analyzed with ion chromatography, are reliable for airborne ammonia measurement.
- The ProTek colorimetric badge system is not recommended due to significant inaccuracies and bias.
- Field tests confirmed good agreement between filter methods, highlighting their suitability for industrial environments.