BM strips: how accurate are they in general wards?
Diabetic Medicine : a Journal of the British Diabetic Association
|September 1, 1984
Summary
BM Test Glycemie 20-800 strips showed significant inaccuracies in general ward plasma glucose testing. Over 32% of results were inaccurate, with 22 cases potentially leading to inappropriate patient management.
Area of Science:
- Clinical Chemistry
- Point-of-Care Testing
- Medical Diagnostics
Background:
- Point-of-care testing (POCT) devices like BM Test Glycemie 20-800 strips are widely used for rapid glucose monitoring.
- Accurate glucose measurements are critical for timely and appropriate patient management, especially in emergency settings.
- General ward settings present unique challenges for maintaining the accuracy of diagnostic tests.
Purpose of the Study:
- To evaluate the accuracy of BM Test Glycemie 20-800 strips for plasma glucose determination in a general ward setting.
- To compare BM strip results with laboratory-based plasma glucose measurements.
Main Methods:
- Simultaneous plasma glucose measurements were performed using BM Test Glycemie 20-800 strips and a Beckman II glucose analyzer.
- A total of 183 emergency out-of-hours plasma glucose requests were analyzed.
- Accuracy was defined as the BM strip value being within one strip-range of the laboratory value.
Main Results:
- Fifty-nine (32.2%) of the BM strip results were deemed inaccurate based on the defined criteria.
- In 22 of the inaccurate results, the discrepancy was significant enough to potentially lead to inappropriate clinical management.
- The specific reasons for the observed inaccuracies were not identified in this study.
Conclusions:
- BM Test Glycemie 20-800 strips demonstrated a concerning level of inaccuracy when used in general wards for plasma glucose testing.
- A significant proportion of inaccurate results could have clinical implications, potentially leading to incorrect treatment decisions.
- Improved training for ward staff and the implementation of formal quality control schemes are recommended to enhance the reliability of these point-of-care glucose tests.