Lateral position improves wedge-left atrial pressure correlation during positive-pressure ventilation
Critical Care Medicine
|November 1, 1984
Summary
The lateral position improves pulmonary artery occlusion pressure (Pw) measurements, offering a more accurate reflection of left atrial pressure (Pla). This technique is beneficial during positive-end expiratory pressure (PEEP) ventilation, but requires further human studies.
Area of Science:
- Cardiovascular Physiology
- Respiratory Physiology
- Medical Instrumentation
Background:
- Pulmonary artery catheterization is crucial for hemodynamic monitoring.
- Accurate measurement of left atrial pressure (Pla) is vital for assessing cardiac function.
- Pulmonary artery occlusion pressure (Pw) is often used as a surrogate for Pla, but its accuracy can be affected by patient positioning and ventilation strategies.
Purpose of the Study:
- To investigate the effect of patient lateral positioning on the correlation between pulmonary artery occlusion pressure (Pw) and left atrial pressure (Pla).
- To evaluate how varying levels of positive-end expiratory pressure (PEEP) influence the accuracy of Pw measurements in different positions.
Main Methods:
- Ten anesthetized pigs were studied using catheters placed in the left atrium and pulmonary arteries.
- Simultaneous measurements of Pla and Pw were recorded in supine, right lateral, and left lateral positions.
- Measurements were taken at various increments of positive-end expiratory pressure (PEEP).
Main Results:
- Pw measurements obtained in the lateral position, with the catheter tip below the left atrium, more accurately reflected Pla changes compared to measurements with the tip above the left atrium, especially at PEEP levels below 20 cm H2O.
- In the supine position, Pw measurements from the lower catheter tip provided a more accurate assessment of Pla than those from the upper tip, particularly when the tip was vertically above the left atrium.
Conclusions:
- The lateral position can enhance the accuracy of pulmonary artery occlusion pressure (Pw) measurements during continuous positive-pressure ventilation.
- This positioning maneuver may improve the reliability of using Pw as a surrogate for left atrial pressure (Pla).
- Further clinical investigations in human subjects are necessary before recommending this technique for routine use.