Related Experiment Videos
[Dosimetric characteristics of the bremsstrahlung beam from the LUE-15M medical linear electron accelerator]
Abstract:
The paper presents methods and results of a study of radiation-physical characteristics of inhibitory radiation beam with the Grenz energy of 15MeV generated by an electron linear accelerator LUE-15M. Special emphasis is laid on primary dosimetric information used for the planning of radiotherapy: depth doses, beam profiles, dose functions of a collimated beam. It has been shown that in general the accelerator meets the requirements of the International Electrotechnical Commission. General error in the focal absorbed dose at the expense of variable parameters of the accelerator was evaluated. It does not exceed +/- 3.5%.