Related Experiment Video
Updated: Sep 24, 2026

Cardiopulmonary Bypass in a Mouse Model: A Novel Approach
Published on: September 22, 2017
Initial clinical experience with a low pressure drop membrane oxygenator for cardiopulmonary bypass in adult patients
Abstract:
The new Travenol oxygenator is composed of 80 parallel blood pathways. Microporous membrane separates the blood and gas compartments. The membrane surface area is 3 m2, with a pore size of 0.01 microns. Venous blood drains directly from the patient through the oxygenator, then through an integral heat exchanger and into a reservoir, from which a single arterial pump returns the blood to the patient. The advantage of this configuration of membrane oxygenator is simplicity of setup and operation. A disadvantage that we have observed is an apparent variation in resistance to blood flow through the oxygenator during clinical perfusion. Construction changes in a later version of the oxygenator have reduced the resistance to flow through the blood pathway. This device has been used for 20 perfusions at moderate hypothermia (mean 31.8 degrees C) in patients up to 2.1 m2 body surface area for up to 313 minutes. Blood flow was 2.1 to 5.6 liters/min, partial arterial oxygen pressure 100 to 394 torr, partial arterial carbon dioxide pressure 19 to 57 torr (mean 37 torr) and, arterial pH 7.29 to 7.56 (mean 7.41). Oxygen transfer was as high as 230 ml/min. This integral oxygenator-heat exchanger-reservoir is operated like a bubble oxygenator, with direct venous drainage through the device and a single pump, but it uses a membrane oxygenator for gas exchange to eliminate the detrimental effects of bubbles.
More Related Videos
Related Concept Videos
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes:
Cardiopulmonary Resuscitation II: ACLS Airway Management