Related Experiment Video
Updated: Aug 15, 2026

X-ray Beam Induced Current Measurements for Multi-Modal X-ray Microscopy of Solar Cells
Published on: August 20, 2019
A statistical analysis of ionometrically measured 6-MV x-ray beam parameters
Abstract:
A statistical analysis has been made of ionometrically measured parameters related to the central-axis output, the beam symmetry, the beam flatness, and the depth dose for the 6-MV x-ray beam from the Therac-6 linear accelerator. The means and coefficients of variation of the measurements are determined on a monthly basis for each parameter. Mean coefficients of variation less than +/- 1% have been determined and are useful in establishing the optimal limits of acceptance of these parameters. The results of this study indicate that the long-term stability of the radiation parameters associated with this unit can be maintained within acceptable limits with a consistent quality-assurance program.

