Investigation of flow rates of oxygen systems used in general aviation
Abstract:
Flow rates of two general aviation constant flow oxygen systems were tested at altitude using a Ted Nelson flow meter and using timed flow into a volume bag. Contrary to the manufacturer's claim, the Rajay SK-10 system was not altitude compensated and provided a flow rate of about 2.2 1/min independent of altitude. The Scott system provided the stated 2.5 l/min with the standard connector and 8 l/min with the "green" connector, the significance of the color code being not widely understood. The conclusions of the 1975 report on oxygen mask performance should be revised according to these flow data. The Ted Nelson flow meter performed within its stated accuracy.
Related Concept Videos
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Pipe Flowrate Measurement
The orifice meter is a simple,...
Pipe Flowrate Measurement: Problem Solving
Bioreactor Controls-II