Inability to control selected drugs on commercially obtained microdilution MIC panels
American Journal of Clinical Pathology
|October 1, 1984
Summary
Many antimicrobial susceptibility testing panels struggle to provide reliable results. This study found that several drugs in Micro Scan panels could not be effectively controlled, impacting the accuracy of minimal inhibitory concentration (MIC) testing.
Area of Science:
- Clinical microbiology
- Antimicrobial resistance
- Diagnostic testing
Background:
- The proliferation of broad-spectrum antimicrobics necessitates comprehensive susceptibility testing.
- Manufacturers often compromise the range of dilutions tested in microdilution panels to accommodate new drugs.
- Accurate minimal inhibitory concentration (MIC) determination is crucial for effective antimicrobial therapy.
Purpose of the Study:
- To evaluate the reliability of Micro Scan 96-well microdilution panels for antimicrobial susceptibility testing.
- To determine if on-scale MIC endpoints could be achieved for drugs in both gram-positive and gram-negative panels.
- To assess the control procedures for specific antimicrobics as per manufacturer instructions.
Main Methods:
- Testing of Micro Scan gram-positive and gram-negative 96-well microdilution panels.
- Use of suggested control organisms over a 60-day period.
- Evaluation of MIC endpoint achievement for individual drugs.
Main Results:
- Seven of 13 antimicrobics in the gram-positive panel showed inadequate control (clindamycin, tetracycline, cephalothin, gentamicin, nitrofurantoin, trimethoprim-sulfa, streptomycin).
- Six of 18 antimicrobics in the gram-negative panel also exhibited poor control (tobramycin, trimethoprim-sulfa, cefoperazone, mezlocillin, ticarcillin, kanamycin).
- The accuracy of patient isolate results for these problematic antimicrobics could not be validated by the provided control procedures.
Conclusions:
- Several key antimicrobics in Micro Scan panels do not meet control criteria for reliable MIC testing.
- The current control procedures may be insufficient to guarantee the accuracy of susceptibility results for certain drugs.
- Further investigation and potential adjustments to panel design or control methods are warranted for improved diagnostic accuracy.