Pressure infusors: variability in delivered infusion pressure.
Critical Care Medicine
|November 1, 1984
Summary
External pressure applied to saline bags is reduced as fluid is withdrawn. This pressure difference increases with decreasing saline volume, impacting continuous flush devices.
Area of Science:
- Medical Devices
- Fluid Dynamics
- Clinical Engineering
Background:
- Pressure infusors are commonly used in clinical settings to deliver fluids.
- Accurate pressure delivery is crucial for various medical procedures and devices.
- The physical properties of fluid bags can influence pressure transmission.
Purpose of the Study:
- To quantify the relationship between applied and delivered pressure from saline bags.
- To investigate how fluid volume affects pressure accuracy in pressure infusion systems.
- To identify potential inaccuracies in pressure delivery for clinical applications.
Main Methods:
- A pressure infusor applied external pressures (300, 250, 200, 150 mm Hg) to 500-ml saline bags.
- Saline was withdrawn in 50-ml aliquots under constant applied pressure.
- Delivered pressure was measured and compared to applied pressure at different volumes.
Main Results:
- Delivered pressure was consistently lower than the applied pressure.
- The discrepancy between applied and delivered pressure increased as the saline volume decreased.
- This pressure drop is an inherent characteristic of the system.
Conclusions:
- The pressure delivered by saline bags is volume-dependent and less than the applied pressure.
- Clinical use, especially with continuous flush devices, must account for this pressure deficit.
- Further consideration is needed for accurate fluid delivery in critical applications.