Safe lateral wall cuff pressure to prevent aspiration.

S Mehta

    Annals of the Royal College of Surgeons of England
    |November 1, 1984
    PubMed
    Summary

    Aspiration remains a risk despite improved tracheal tubes. This study measured distances in supine adults to suggest an optimal intracuff pressure of 2.5 to 3kPa for effective tracheal sealing and reduced aspiration risk.

    Related Concept Videos

    You might also read

    Related Articles

    Articles linked to this work by shared authors, journal, and citation graph.

    Sort by
    Same author

    Ready access to sialylated oligosaccharide donors.

    Organic letters·2001
    Same author

    Preparation of 1,5-disubstituted pyrrolidin-2-ones.

    The Journal of organic chemistry·2000
    Same author

    Prediction of post-extubation work of breathing.

    Critical care medicine·2000
    Same author

    Immunosuppressive activity in buffalo placenta.

    Indian journal of experimental biology·2000
    Same author

    Non-invasive ventilation in acute hypoxemic respiratory failure.

    Intensive care medicine·2000
    Same author

    Control of tracheal cuff pressure.

    Anaesthesia·2000

    Area of Science:

    • Anesthesiology
    • Critical Care Medicine
    • Otolaryngology

    Background:

    • Aspiration is a significant risk during mechanical ventilation, even with modern large-volume, low-pressure endotracheal tube cuffs.
    • Tracheal damage from cuffs has decreased, but aspiration remains a persistent challenge in patient care.

    Purpose of the Study:

    • To determine the maximum hydrostatic pressure that can be generated by a fluid column above the tracheal cuff in supine adult patients.
    • To establish an optimal intracuff pressure for endotracheal tubes to ensure an effective tracheal seal and minimize aspiration risk.

    Main Methods:

    • Measurements of vertical and horizontal distances between the upper central incisor teeth and the suprasternal notch were taken in 200 adult subjects in a supine position.
    • Analysis of anatomical distances to infer potential hydrostatic pressures within the airway.

    Main Results:

    • The maximum vertical distance measured was 9 cm.
    • The maximum horizontal distance measured was 21 cm.
    • These measurements provide a basis for calculating potential hydrostatic pressures above the tracheal cuff.

    Conclusions:

    • A fixed intracuff pressure between 2.5 to 3kPa is recommended for endotracheal tubes.
    • This pressure range is suggested to achieve an optimal tracheal seal, thereby potentially reducing the incidence of aspiration.
    • Findings have implications for airway management protocols and patient safety during mechanical ventilation.