Evidence for surface effects in an electron-capture detector.

A Nazareth, K O'Connell, A Sentissi

    Journal of Chromatography
    |November 30, 1984
    PubMed
    Summary

    Surface effects in electron-capture detectors were reduced in a new inert version. This inert detector shows more consistent solute response patterns for pesticides and electrophores.

    Related Concept Videos

    You might also read

    Related Articles

    Articles linked to this work by shared authors, journal, and citation graph.

    Sort by
    Same author

    A decade of Clostridioides difficile infection: a constant challenge to maintain the status quo.

    The Journal of hospital infection·2023
    Same author

    Antimicrobial overuse in COVID-19 - reasons to remain vigilant as we approach the winter 'twindemic'.

    The Journal of hospital infection·2022
    Same author

    Multidisciplinary neurosurgical rounds incorporating antimicrobial stewardship. Are they of benefit?

    Brain & spine·2022
    Same author

    The early test catches the case. Why wait? Frequent testing of close contacts aids COVID-19 control.

    The Journal of hospital infection·2021
    Same author

    Evolving epidemiology of carbapenemase-producing Enterobacterales: one hospital's infection prevention and control response over nine years.

    The Journal of hospital infection·2021
    Same author

    Casting the net wide: universal testing of emergency admissions for SARS-CoV-2 to prevent onward transmission.

    The Journal of hospital infection·2020

    Area of Science:

    • Analytical Chemistry
    • Chromatography

    Background:

    • Electron-capture detectors (ECDs) are widely used in chromatography.
    • Surface effects in conventional ECDs can lead to non-ideal solute response.
    • Varian developed an experimental, more inert ECD to mitigate these issues.

    Purpose of the Study:

    • To investigate and characterize surface effects in conventional and inert electron-capture detectors.
    • To evaluate the impact of these surface effects on the response of various solutes.
    • To demonstrate the improved performance of the inert ECD.

    Main Methods:

    • Comparison of a conventional ECD with an experimental inert ECD.
    • Analysis of pesticides (herbicides) and derivatized cytosine electrophores.
    • Concentration-dependent response factor analysis using lindane as a model compound.

    Main Results:

    • Surface effects were significantly reduced in the inert ECD compared to the conventional version.
    • Unique solute response patterns observed with the conventional ECD were mitigated in the inert version.
    • The response factor exhibited a minimum followed by a pronounced maximum with increasing solute concentration in the conventional ECD, an effect lessened in the inert version.

    Conclusions:

    • The experimental inert ECD demonstrates a significant reduction in problematic surface effects.
    • The inert ECD offers more reliable and consistent solute response, particularly for electrophilic compounds.
    • Further investigation into the mechanisms underlying these surface effects is warranted.