Interactions between hearing aid and patient in the individual fitting situation
Scandinavian Audiology
|January 1, 1984
Summary
Hearing aid fitting varies due to individual ear canal differences. A universal adjustment factor is not feasible because hearing aid sensitivity to acoustic load impedance varies significantly between patients.
Area of Science:
- Audiology
- Acoustics
- Biomedical Engineering
Background:
- Individual anatomical variations in the ear canal and middle ear affect hearing aid performance.
- Standardized measurements using KEMAR (Knowles Electronics Manikin for Acoustic Research) provide a baseline but do not fully account for patient-specific acoustics.
Purpose of the Study:
- To investigate the feasibility of introducing an adjustment factor for hearing aid fitting based on individual ear mould characteristics.
- To quantify the variation in insertion gain across different hearing aids when fitted to the same individual.
Main Methods:
- Testing multiple hearing aids on the same individual with custom ear moulds.
- Comparing the measured insertion gain against KEMAR data to assess deviations.
Main Results:
- Significant variations in insertion gain were observed for identical hearing aids across different individuals and even for the same individual with different hearing aids.
- The extent of variation indicated that a single, universal adjustment factor is not directly applicable.
Conclusions:
- Hearing aid fitting requires consideration of individual acoustic load impedance.
- Predicting hearing aid effectiveness necessitates understanding the device's sensitivity to specific patient ear canal acoustics.