Radiation shielding in dental radiography.
Summary
Radiation shielding effectively reduces radiation dose to the thyroid, sternal, and gonadal regions during dental radiography. Different leaded aprons and collars offer significant protection, with a horizontal shield providing the most substantial dose reduction.
Area of Science:
- Radiological Protection
- Medical Physics
- Dental Imaging
Background:
- Intraoral radiography utilizes ionizing radiation, necessitating protective measures for patients.
- Assessing the efficacy of various radiation shielding devices is crucial for minimizing patient dose.
Purpose of the Study:
- To evaluate the protective effect of different radiation shielding types on the thyroid, sternal, and gonadal regions during intraoral radiography.
- To quantify the absorbed dose reduction achieved by various shielding configurations.
Main Methods:
- Tested leaded aprons with attached or separate collars, and a horizontal leaded shield.
- Measured absorbed doses in the thyroid, sternal, and gonadal regions using ionization chambers.
- Utilized two dental X-ray machines (60 and 65 kVp) with different collimators and exposure techniques.
Main Results:
- No significant difference in thyroid protection was found between different shielding types.
- Thyroid dose was reduced by approximately a factor of 2-5, depending on the technique.
- Sternal dose was reduced to below 2 microGy with apron and collar combinations, and by a factor of 6 with a horizontal shield.
- Gonadal dose was reduced by a factor of 10 with the horizontal shield, paralleling technique, and circular collimator.
Conclusions:
- Various radiation shielding devices provide significant dose reduction to critical organs during intraoral radiography.
- Leaded aprons, collars, and horizontal shields are effective in minimizing patient radiation exposure.
- The choice of shielding and technique impacts the degree of dose reduction achieved.