Calibration Procedures for Orthogonal Superposition Rheology
Published on: November 18, 2020
Improving accuracy of co-axial viscometry
Abstract:
Aggregation and sedimentation of the red cells in the viscometer makes it difficult to obtain good reproducibility at low shear rates. A simple method is described to obtain repeatable torque-time curves with the Contraves LS30. Before starting rotation, the blood sample is agitated by an up and down movement of the bob. This results in identical (within 1-2% of the actual amplitude) repeated tracings even at the lowest shear rates, even with extremely time-dependent samples. The subsequent torque-time curve of the disaggregated and unsettled samples provides an experimental basis for the quantitative description of blood viscoelasticity and for dynamic analysis of red cell aggregation: furthermore it results in more accurate measurements when comparing different samples and testing therapeutic effects. The method is applicable to any coaxial cylinder type viscometer with manual lowering of the bob.
Related Concept Videos
Uncertainty in Measurement: Accuracy and Precision
Uncertainty in Measurement: Reading Instruments
Accuracy and Precision
Accuracy, limits, and approximation
Accuracy is defined as the closeness of the measured value to the true or actual value. In engineering mechanics, repeated measurements are taken during theoretical or experimental analyses to ensure that the result is precise and accurate.
The accuracy of any solution is based on the...
Contact Angle
The adhesive force is the molecular force between molecules of different materials, that is, between the molecules of the solid and the liquid. The cohesive force...
Accuracy and Precision