Jove
Visualize
Contact Us

Related Experiment Videos

Neonatal magnification radiography using standard-focus x-ray tubes.

W Nikesch, C M Kuntzler, F R Cushing

    AJR. American Journal of Roentgenology
    |October 1, 1983
    PubMed
    Summary

    This study proposes using standard mobile X-ray units for moderate magnification radiography in premature infants. This technique improves image quality by reducing noise and enhancing contrast, making radiographic examinations clearer.

    Related Concept Videos

    You might also read

    Related Articles

    Articles linked to this work by shared authors, journal, and citation graph.

    Sort by
    Same author

    False left ventricular aneurysm with ventriculo-bronchial fistula and massive haemoptysis.

    Scandinavian cardiovascular journal : SCJ·1997
    Same author

    Bacteremia and hemoptysis due to an infected false left ventricular aneurysm associated with a bronchoventricular fistula.

    Clinical infectious diseases : an official publication of the Infectious Diseases Society of America·1996
    Same author

    Three-dimensional computerized tomography: a quantum leap in diagnostic imaging?

    The Journal of foot and ankle surgery : official publication of the American College of Foot and Ankle Surgeons·1994
    Same author

    Problems with the Radiation Measurements Incorporated/American College of Radiology mammography accreditation phantom.

    Medical physics·1990
    Same author

    Use of spreadsheets to manipulate field scanning data.

    Medical dosimetry : official journal of the American Association of Medical Dosimetrists·1990
    Same author

    Exposure reduction using yttrium filters in a cardiac catheterization unit.

    Catheterization and cardiovascular diagnosis·1986
    JoVE
    x logofacebook logolinkedin logoyoutube logo
    ABOUT JoVE
    OverviewLeadershipBlogJoVE Help Center
    AUTHORS
    Publishing ProcessEditorial BoardScope & PoliciesPeer ReviewFAQSubmit
    LIBRARIANS
    TestimonialsSubscriptionsAccessResourcesLibrary Advisory BoardFAQ
    RESEARCH
    JoVE JournalMethods CollectionsJoVE Encyclopedia of ExperimentsArchive
    EDUCATION
    JoVE CoreJoVE BusinessJoVE Science EducationJoVE Lab ManualFaculty Resource CenterFaculty Site
    Terms & Conditions of Use
    Privacy Policy
    Policies

    Area of Science:

    • Medical Imaging
    • Radiography
    • Neonatal Care

    Background:

    • Radiographic examination of premature infants is challenging due to small size and low contrast.
    • Magnification radiography can enhance image quality but typically requires specialized microfocus equipment.
    • Such equipment is often unavailable in neonatal intensive care units.

    Purpose of the Study:

    • To propose a cost-effective method for moderate magnification radiography (1.5x) using standard mobile X-ray units.
    • To evaluate the feasibility of using the anode side of the X-ray field for improved image quality in neonatal radiography.
    • To demonstrate that standard equipment can achieve superior radiographic results in premature infants.

    Main Methods:

    • Utilizing standard-focus X-ray tubes (≤1 mm focal spot) with mobile units.

    Related Experiment Videos

  • Employing the anode side of the X-ray field for moderate magnification.
  • Leveraging high tube current for short exposure times and long target-film distances.
  • Implementing a large air gap to improve image contrast.
  • Main Results:

    • Magnification radiographs produced using the anode side were superior to those from the central beam and standard contact radiographs.
    • The technique resulted in decreased noise and improved contrast in the magnified images.
    • Short exposure times and long target-film distances minimized motion unsharpness.

    Conclusions:

    • Moderate magnification radiography with standard mobile X-ray units is a viable and cost-effective alternative for neonatal imaging.
    • This method significantly improves image quality, aiding in the radiographic examination of premature infants.
    • The proposed technique overcomes the limitations of specialized equipment availability in neonatal intensive care units.