Thermodilution method overestimates low cardiac output in humans
The American Journal of Physiology
|October 11, 1983
Summary
The thermodilution method may overestimate cardiac output in low-flow states, potentially due to heat loss. This finding is crucial for accurately assessing patients with low cardiac output.
Area of Science:
- Cardiovascular Physiology
- Medical Device Technology
Background:
- Cardiac output (CO) measurement is vital for managing critically ill patients.
- Thermodilution and Fick methods are commonly used to assess CO.
- Previous studies have shown discrepancies between CO measurement techniques.
Purpose of the Study:
- To compare cardiac output measurements between thermodilution and Fick methods.
- To investigate potential overestimation of cardiac output by thermodilution in low-flow states.
Main Methods:
- Compared 57 cardiac output measurements using thermodilution and Fick methods in 26 patients.
- Analyzed discrepancies in measurements, particularly in low output ranges (Fick < 3.5 L/min).
Main Results:
- Thermodilution values were higher than Fick values in all 16 cases with Fick outputs < 3.5 L/min.
- An average difference of 35% was observed when Fick values were ≤ 2.5 L/min.
- Findings suggest thermodilution overestimates true cardiac output in low output ranges, likely due to heat loss.
Conclusions:
- The thermodilution method may overestimate cardiac output in low-flow conditions.
- Heat loss is a probable cause for this overestimation.
- Clinical implications exist for patients with low output states where thermodilution is frequently used.