Blood collection while using a continuous glucose analyzer without insertion of an additional venous catheter
Diabetologia
|August 1, 1983
Summary
A novel method enables continuous blood collection for glucose monitoring using the Biostator. This technique allows excess blood collection for assays without compromising glucose measurement accuracy.
Area of Science:
- Biomedical Engineering
- Clinical Chemistry
Background:
- Continuous glucose monitoring is crucial for diabetes management.
- Existing methods may have limitations in blood sample handling for additional assays.
Purpose of the Study:
- To describe a modified Biostator method for continuous blood collection.
- To evaluate the impact of this method on blood glucose measurements and sample yield.
Main Methods:
- Utilizing a double lumen catheter and an infusion pump's optional channel for blood withdrawal.
- Collecting excess blood into assay tubes for secondary analysis.
- Calculating a dilution factor based on tubing diameter to correct for sample dilution.
Main Results:
- Continuous collection of diluted heparinized blood was achieved.
- A mean volume of 1.21 +/- 0.07 ml/10 min was collected using a 0.020-inch lumen diameter tube.
- The method showed no significant impact on blood glucose measurement accuracy (r = 0.9572).
Conclusions:
- The described modification provides a reliable method for continuous blood collection alongside glucose monitoring.
- This technique allows for additional blood assays without compromising the primary glucose measurement.
- Further research may involve optimizing dilution correction factors for various experimental conditions.