A scanning system for chest radiography with regional exposure control: practical implementation.

D B Plewes, E Vogelstein

    Medical Physics
    |September 1, 1983
    PubMed
    Summary

    This study introduces a novel scanning apparatus for chest radiography, improving image quality through superior scatter rejection and exposure control. The system enhances soft tissue contrast and uniformity, particularly in challenging anatomical regions.

    Related Concept Videos

    You might also read

    Related Articles

    Articles linked to this work by shared authors, journal, and citation graph.

    Sort by
    Same author

    Magnetic monitoring of a small Foucault pendulum.

    The Review of scientific instruments·2018
    Same author

    Long-term results of screening with magnetic resonance imaging in women with BRCA mutations.

    British journal of cancer·2012
    Same author

    Flow-induced disturbances in balanced steady-state free precession images: means to reduce or exploit them.

    Magnetic resonance in medicine·2009
    Same author

    Removing undersampling artifacts in DCE-MRI studies using independent components analysis.

    Magnetic resonance in medicine·2008
    Same author

    Tissue ultrasound absorption measurement with MRI calorimetry.

    IEEE transactions on ultrasonics, ferroelectrics, and frequency control·2008
    Same author

    Evaluating an optical-flow-based registration algorithm for contrast-enhanced magnetic resonance imaging of the breast.

    Physics in medicine and biology·2007

    Area of Science:

    • Medical Imaging
    • Radiography Technology
    • Diagnostic Radiology

    Background:

    • Conventional chest radiography faces challenges with scatter radiation and achieving uniform film exposure.
    • Improving soft tissue contrast in mediastinal, diaphragmatic, and retrocardiac areas remains a clinical goal.

    Purpose of the Study:

    • To describe and evaluate an experimental scanning apparatus for chest radiography.
    • To assess the scatter rejection capabilities and exposure control of the new system.
    • To demonstrate improvements in image uniformity and soft tissue contrast.

    Main Methods:

    • An experimental scanning apparatus utilizing two geometries: lateral fan beam sweep (5s) and 2D raster scan (8.8s).
    • Pulse width modulation for tube output control, with film exposure monitored by a fluorescence detector.

    Related Experiment Videos

  • Scatter rejection measurements compared to a 10:1 grid at 120-kVp.
  • Phantom and clinical imaging to evaluate uniformity, contrast, and motion unsharpness.
  • Main Results:

    • The scanning apparatus demonstrated scatter rejection superior to a 10:1 grid at 120-kVp.
    • Phantom images showed improved uniformity and soft tissue contrast in key thoracic areas.
    • Clinical images revealed motion unsharpness at 8.8s, which was reduced at 4.5s scan time.

    Conclusions:

    • The experimental scanning apparatus offers significant advantages in scatter rejection and exposure control for chest radiography.
    • The system has the potential to enhance diagnostic image quality by improving uniformity and soft tissue contrast.
    • Scan time optimization is crucial for mitigating motion unsharpness in clinical applications.