Related Experiment Videos
Ultrasonic attenuation and absorption in liver tissue.
Ultrasound in Medicine & Biology
|July 1, 1983
Summary
Ultrasonic absorption and attenuation in bovine liver are similar at low megahertz frequencies. Absorption is the primary cause of ultrasonic signal attenuation in liver tissue within this range.
Area of Science:
- Biomedical Ultrasound
- Acoustics
- Tissue Optics
Background:
- Discrepancies exist in reported ultrasonic attenuation and absorption coefficients for biological tissues.
- Distinguishing true ultrasonic absorption (energy conversion to heat) from total attenuation is crucial for theoretical and practical applications.
Purpose of the Study:
- To investigate and quantify ultrasonic attenuation and absorption coefficients in bovine liver.
- To determine the relative contributions of absorption and other attenuation mechanisms in liver tissue.
Main Methods:
- Total ultrasonic attenuation was measured in bovine liver across the 1-6 MHz frequency range using phase-sensitive and phase-insensitive insertion loss techniques.
- Ultrasonic absorption coefficients were determined using two thermal methods: the transient thermoelectric (rate-of-heating) method and a novel temperature decay measurement post-ultrasonic pulse.
Main Results:
- Ultrasonic amplitude attenuation and absorption coefficients in bovine liver were found to be not significantly different at low megahertz frequencies.
- Mean values for both coefficients clustered around 0.05 nepers/cm/MHz (0.4 dB/cm/MHz).
- Sample-to-sample variation was minimal, with a standard deviation of 0.01 nepers/cm/MHz (0.09 dB/cm/MHz) or less.
Conclusions:
- In the studied frequency range (1-6 MHz), ultrasonic absorption is the dominant factor contributing to the total attenuation in liver tissue.
- The findings suggest that absorption accounts for most of the ultrasonic signal loss in liver within this frequency spectrum.