Stress-protection after external fixation on the intact rabbit tibia
Acta Orthopaedica Scandinavica
|August 1, 1983
Summary
External fixation on rabbit tibias caused bone loss after 12 weeks, indicating a stress-protecting effect. This bone weakening was less severe than that seen with metal plates.
Area of Science:
- Orthopedics
- Biomedical Engineering
- Bone Physiology
Background:
- Metal plates cause bone loss due to stress protection.
- External fixation is used in bone fracture treatment.
Purpose of the Study:
- To investigate if external fixation causes a stress-protecting effect on intact bone.
- To compare the effect of external fixation to that of metal plates.
Main Methods:
- External mini-fixation applied to rabbit tibiae.
- Biomechanical testing (three-point bending) and bone mineral content measurement (photon absorptiometry) at 6 and 12 weeks.
Main Results:
- No significant changes at 6 weeks.
- Significant reduction in bone strength (87%) and stiffness (88%) at 12 weeks.
- Significant reduction in bone mineral content (90%) in the stabilized segment at 12 weeks.
Conclusions:
- External mini-fixation induces a stress-protecting effect on rabbit tibiae.
- This effect occurs later and is less pronounced than with metal plates.