Related Experiment Video
Updated: Sep 3, 2026

Vascular Balloon Injury and Intraluminal Administration in Rat Carotid Artery
Published on: December 23, 2014
Embolization of small vessels with a double-lumen microballoon catheter. Part II: Laboratory, animal, and
Abstract:
A double-lumen microballoon catheter system measuring 0.5-0.85 mm in outer diameter (1.3-2.8 F) was fabricated for passage through standard 4.1 and 5 F catheters. By arresting circulation with the balloon, embolization with low-viscosity silicone rubber, ethanol, or isobutyl-2-cyanoacrylate can proceed in a methodical, unhurried manner, permitting precise control of tissue penetration and preventing paradoxical embolization of other vessels. Problems such as gluing the catheter in situ with cyanoacrylate did not occur. The catheter may also be inserted during surgery to control embolization at a remote site, which did result in the catheter being glued in situ in one case.