Autoclave sterilization of polypropylene plasma separator
Artificial Organs
|February 1, 1984
Summary
A new water-prefilled plasma separator (PS) sterilized by autoclaving maintains its filtration and sieving properties. This development improves handling and suggests potential clinical use for the plasma separator.
Area of Science:
- Biomedical Engineering
- Materials Science
Background:
- Autoclave sterilization is a common method for medical devices.
- Plasma separators (PS) require careful handling and sterilization.
- Improving pre-use handling procedures for PS is crucial for clinical application.
Purpose of the Study:
- To develop a water-prefilled, autoclave-sterilized plasma separator (PS).
- To evaluate the effect of autoclaving on polypropylene membranes with different pore sizes.
- To assess the filtration and sieving performance of the sterilized PS.
Main Methods:
- Development of a water-prefilled plasma separator (PS).
- Autoclave sterilization of PS with two types of polypropylene membranes (varying pore sizes).
- In vitro filtration study using bovine blood to evaluate performance.
Main Results:
- Larger pore size membranes exhibited significantly higher filtration rates (p < 0.05).
- Sieving coefficients for proteins were comparable between membranes of different pore sizes.
- Water-prefilled, autoclave-sterilized PS demonstrated filtration and sieving properties similar to non-autoclaved PS.
Conclusions:
- Autoclave sterilization of water-prefilled plasma separators does not significantly alter their performance.
- The developed plasma separator shows promising filtration and sieving characteristics.
- The findings suggest the potential clinical acceptability of the autoclave-sterilized plasma separator.