Flow injection analysis-atomic absorption determination of serum zinc
Clinica Chimica Acta; International Journal of Clinical Chemistry
|February 28, 1984
Summary
Flow injection analysis-atomic absorption (FIA-AA) offers a rapid method for determining zinc in human serum. This technique provides comparable results to conventional atomic absorption spectrometry, enabling high-throughput analysis.
Area of Science:
- Analytical Chemistry
- Clinical Chemistry
Background:
- Accurate zinc determination in human serum is crucial for diagnosing and monitoring various health conditions.
- Conventional atomic absorption spectrometry (AAS) is a standard method but can be time-consuming for high-throughput screening.
Purpose of the Study:
- To demonstrate the efficacy of flow injection analysis-atomic absorption (FIA-AA) for quantifying zinc in human serum.
- To compare the results obtained by FIA-AA with those from conventional AAS.
- To optimize FIA-AA parameters for efficient serum zinc analysis.
Main Methods:
- Human serum samples were analyzed for zinc using a flow injection analysis system coupled with atomic absorption spectrometry.
- Protein precipitation was performed using 25% trichloroacetic acid, and 100-microliter aliquots of the filtrate were injected.
- The system utilized the nebulizer's negative pressure for carrier stream aspiration, eliminating the need for a pump.
- The impact of sample volume and dispersion coil length on the FIA signal was investigated.
Main Results:
- FIA-AA provided zinc determination results comparable to conventional AAS, showing no significant difference.
- The method achieved a precision of +/-3.47% relative standard deviation.
- The system demonstrated a high measurement frequency of up to 180 samples per hour.
- A simple 1:5 dilution with water followed by FIA-AA measurements yielded higher zinc levels, suggesting potential matrix effects or dilution benefits.
Conclusions:
- Flow injection analysis-atomic absorption is a viable and efficient alternative for determining zinc in human serum.
- The FIA-AA method is accurate, precise, and capable of high-throughput analysis, making it suitable for clinical laboratories.
- Further studies on sample preparation and FIA parameters can optimize zinc quantification in biological samples.