A probe earmold system for measuring eardrum SPL under hearing-aid conditions
Scandinavian Audiology
|January 1, 1984
Summary
A new probe-earmold (PE) system accurately measures sound pressure levels (SPL) at the eardrum in hearing-aid users. This validated system provides reliable eardrum SPL measurements up to 6.0 kHz in occluded ears.
Area of Science:
- Audiology
- Acoustics
- Biomedical Engineering
Background:
- Accurate measurement of sound pressure level (SPL) at the eardrum is crucial for hearing aid fitting.
- Existing methods may be invasive or inaccurate in occluded ear conditions.
Purpose of the Study:
- To develop and validate a miniature probe-microphone assembly integrated with an earmold (PE system) for measuring eardrum SPL.
- To assess the accuracy of the PE system in human ears under hearing-aid conditions.
Main Methods:
- A probe-earmold (PE) system was designed, integrating the probe with the earmold to avoid ear canal intrusion.
- Previous validation was performed using an ear simulator, showing +/- 3 dB accuracy up to 6.0 kHz.
- A behavioral experiment on human ears compared pure-tone thresholds from a conventional earphone to those from the PE system with a hearing aid receiver.
Main Results:
- The PE system demonstrated accurate eardrum SPL measurements up to 6.0 kHz in an ear simulator.
- No statistically significant differences were found between pure-tone thresholds measured by the conventional earphone and the PE system on human ears.
- These findings confirm the simulator data and validate the PE system's accuracy in occluded ears.
Conclusions:
- The developed probe-earmold (PE) system accurately measures sound pressure level (SPL) at the eardrum in occluded ears up to 6.0 kHz.
- The PE system offers a reliable tool for audiological assessments and hearing aid fitting.
- Validation on human ears confirms the utility of the PE system for clinical applications.