Characterization of chronically implanted ventricular endocardial pacing leads.

D A Angello, J H McAnulty, J Dobbs

    American Heart Journal
    |June 1, 1984
    PubMed
    Summary

    Pacemaker lead integrity is crucial as generator life extends. This study analyzed chronic lead data, revealing thresholds and impedances that guide long-term pacemaker assessments.

    Related Concept Videos

    You might also read

    Related Articles

    Articles linked to this work by shared authors, journal, and citation graph.

    Sort by
    Same author

    United Kingdom Radiation Oncology 1 Conference (UKRO 1): accuracy and uncertainty in radiotherapy.

    The British journal of radiology·2002
    Same author

    Prevalence of tics and Tourette syndrome in an inpatient adult psychiatry setting.

    Journal of psychiatry & neuroscience : JPN·2002
    Same author

    Carotid sinus massage in patients who fall: will it define the role of pacing?

    Journal of the American College of Cardiology·2001
    Same author

    Patients at lower risk of arrhythmia recurrence: a subgroup in whom implantable defibrillators may not offer benefit. Antiarrhythmics Versus Implantable Defibrillator (AVID) Trial Investigators.

    Journal of the American College of Cardiology·2001
    Same author

    A comparison of Medicare fee-for-service and a group-model HMO in the inpatient management and long-term survival of elderly individuals with syncope.

    The American journal of managed care·2001
    Same author

    "Stable" ventricular tachycardia is not a benign rhythm : insights from the antiarrhythmics versus implantable defibrillators (AVID) registry.

    Circulation·2001

    Area of Science:

    • Cardiology
    • Biomedical Engineering
    • Medical Device Technology

    Background:

    • Advancements in lithium power sources have increased pacemaker generator longevity.
    • This shift makes the pacing lead the limiting factor for overall pacemaker implant lifespan.
    • Accurate assessment of chronically implanted pacing lead integrity is therefore essential.

    Purpose of the Study:

    • To retrospectively analyze chronic threshold and impedance data from ventricular endocardial leads.
    • To establish general guidelines for chronic pacing lead parameters based on long-term data.
    • To highlight challenges in collecting and interpreting long-term pacing lead performance data.

    Main Methods:

    • Retrospective analysis of chronic threshold and impedance data.

    Related Experiment Videos

  • Inclusion of 61 patients with the same type of ventricular endocardial lead.
  • Data collection spanned up to 10 years of chronic implantation.
  • Main Results:

    • Mean acute current threshold was 0.81 +/- 0.33 mA.
    • Chronic current thresholds ranged from 1 to 5 mA, 2-5 times the acute implant value.
    • Chronic voltage thresholds ranged from 0.7 to 4 V, and impedance from 335 to 775 omega.

    Conclusions:

    • The study provides general guidelines for interpreting chronic pacing lead parameters.
    • Findings underscore the significant increase in chronic thresholds compared to acute values.
    • Interpreting long-term lead data presents notable difficulties that require careful consideration.