Measuring the Carotid to Femoral Pulse Wave Velocity (Cf-PWV) to Evaluate Arterial Stiffness
Published on: May 3, 2018
Indirect blood pressure measurement at the common femoral artery
Insights
This study validates a novel indirect method for measuring common femoral artery blood pressure using a long thigh cuff and Doppler flow signals. The technique proved reliable and reproducible, offering a valuable non-invasive option for blood pressure assessment.
Area of Science:
- Vascular Medicine
- Cardiovascular Diagnostics
- Biomedical Engineering
Background:
- Indirect blood pressure measurement is crucial for cardiovascular health assessment.
- Assessing blood pressure at the common femoral artery level non-invasively presents unique challenges.
- Existing methods may lack accuracy or applicability in certain patient populations.
Purpose of the Study:
- To describe and evaluate the reliability of a new indirect method for measuring blood pressure at the common femoral artery.
- To assess the correlation between indirect femoral cuff pressures and direct intra-arterial measurements.
- To determine the reproducibility of this indirect measurement technique.
Main Methods:
- Utilized a long thigh cuff encircling the hips, with a pediatric blood pressure cuff placed over the common femoral artery to transmit pressure.
- Doppler technique applied at the ankle determined flow signals.
- Cuff bladder widths of 14 cm or 16 cm were used based on patient hip circumference.
- Measurements were validated against common femoral intra-arterial pressures in 24 subjects.
Main Results:
- A close correlation was observed between indirect femoral cuff pressures and direct common femoral intra-arterial pressures.
- The indirect measurement technique demonstrated similar reproducibility to measurements at the ankle and toe levels.
- Successful measurements were achieved even in patients with iliac disease and collateral vessels.
Conclusions:
- The described indirect method for measuring common femoral artery blood pressure is reliable and reproducible.
- This non-invasive technique offers a viable alternative for assessing peripheral blood pressure, particularly in challenging cases.
- The findings support the use of this technique in clinical practice for enhanced cardiovascular assessment.
Abstract:
In the present study the technique and the reliability of indirect blood pressure measurement at the common femoral level was described. The blood pressure was measured by using a long cuff, which encircled the hips at the level of the groin, and determining the flow signals by Doppler technique applied at the ankle. In order to transmit the cuff pressure to the artery, a slender bag with air, which was a blood pressure cuff used in children, was placed over the common femoral artery under the cuff. A 14 cm. wide cuff-bladder was usually used, but in obese subjects with the size of hips more than 90 cm, a 16 cm. wide cuff-bladder was applied. In patients with collateral vessels around the groin due to the iliac disease, the blood pressure could be measured by placing the Doppler flat probe over the common femoral artery or the collateral vessel. In observation of 24 subjects, a close correlation was obtained between the femoral cuff pressures and common femoral intra-arterial pressures. Furthermore, blood pressure values at the common femoral level measured indirectly showed similar reproducibility as those at the ankle and toe levels.
Related Concept Videos
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(two-step method)
Assessing Blood pressure in the Leg
Preparation:
Pre-Procedural Guidelines for Assessing Blood Pressure
Measurement of Blood Pressure