Sample viscosity can be a source of analytical error when discrete sampler-dilutors are used
Clinical Chemistry
|November 1, 1981
Summary
Hyperviscosity syndrome can cause inaccurate total protein measurements using automated systems like the DuPont aca. Using a manual syringe method for sample addition to the aca eliminates these viscosity-dependent errors.
Area of Science:
- Clinical Chemistry
- Biomedical Engineering
- Laboratory Medicine
Background:
- Hyperviscosity syndrome presents diagnostic challenges due to altered physiological properties.
- Accurate protein quantification is crucial for patient management.
- Automated laboratory analyzers may exhibit performance variations with non-ideal sample matrices.
Purpose of the Study:
- To investigate the discrepancy in total protein measurements for a hyperviscosity syndrome patient.
- To identify the cause of viscosity-dependent errors in the DuPont aca analyzer.
- To propose a method for mitigating these analytical errors.
Main Methods:
- Comparative analysis of total protein concentration using the biuret method on DuPont aca and SMA 12/60 analyzers.
- Evaluation of the impact of sample temperature and aspirated volume on aca measurements.
- Assessment of alternative sampling methods, including manual syringe addition and Micromedic sampler-dilutor.
Main Results:
- The DuPont aca reported significantly lower total protein (80 g/L) compared to the SMA 12/60 (105 g/L) for the patient's serum.
- Viscosity-dependent errors in the aca were confirmed, influenced by sample temperature and volume.
- Error magnitude in the aca was reduced with a Micromedic sampler-dilutor and eliminated by manual syringe addition.
Conclusions:
- Automated sampling systems, particularly the DuPont aca, are susceptible to viscosity-dependent errors in protein assays.
- Sample viscosity significantly impacts the accuracy of automated clinical analyzers.
- Manual syringe addition is recommended for analyzing highly viscous samples (e.g., serum in hyperviscosity syndrome) with the DuPont aca to ensure accurate results.