Abstract:
A quantitative method for determination of nitroglycerin in human plasma was developed. Nitroglycerin and the internal standard (butane-1,2,4-triyl trinitrate) were extracted from plasma with pentane. The extracts were analysed by gas chromatography mass spectrometry using fused silica capillary columns and electron capture negative ion chemical ionization. The quantitation limit of the method was about 50 pg ml-1. Linear calibration curves were obtained in the range of 50-1600 pg ml-1. Precision at the level of 100 pg ml-1 was 4%.
Related Concept Videos
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall. The coating...
Precipitation Gravimetry
In determining nickel by gravimetric analysis, a precipitant of ethanolic dimethylglyoxime is added to a hot nickel salt solution. This is quickly followed by the dropwise addition of dilute ammonia solution until precipitation occurs. A...
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Gas Chromatography: Types of Detectors-II