Related Experiment Video
Updated: Aug 12, 2026

Laser-heating and Radiance Spectrometry for the Study of Nuclear Materials in Conditions Simulating a Nuclear Power Plant Accident
Published on: December 14, 2017
Unwanted radiation produced by leakage neutrons from medical electron accelerators
Abstract:
Monte Carlo calculations have been made to determine the energy delivered to a phantom by neutrons escaping from the head of a Varian Clinac 35 medical accelerator at an SSD of 1 m. The energy was sorted into two regions: inside and outside a volume defined by a circular beam of area 100 and 600 cm2. For the two beam sizes, the energies outside the treatment volume were 12 and 8.7 g-rad (per photon rad) respectively. Room scattering increased these values by about 20%. These energies are smaller than those delivered by scattered photons by more than ten times.