Performance of the Hitachi 705 evaluated.

P Douville, J C Forest

    Clinical Chemistry
    |April 1, 1983
    PubMed
    Summary

    The Hitachi 705 discrete analyzer offers high throughput for 19 analytes. It demonstrated good reliability and practicability for most clinical chemistry assays, though calcium and phosphorus require further review.

    Related Concept Videos

    You might also read

    Related Articles

    Articles linked to this work by shared authors, journal, and citation graph.

    Sort by
    Same author

    Alteration of calcium homeostasis in primary preeclamptic syncytiotrophoblasts: effect on calcium exchange in placenta.

    Journal of cellular and molecular medicine·2010
    Same author

    Standardization in laboratory medicine: new challenges.

    Clinica chimica acta; international journal of clinical chemistry·2005
    Same author

    Estrogen/progesterone replacement versus pravastatin and their sequential association in hypercholesterolemic postmenopausal women.

    Maturitas·2001
    Same author

    Effects of electric and magnetic fields from high-power lines on female urinary excretion of 6-sulfatoxymelatonin.

    American journal of epidemiology·2001
    Same author

    Quantification of all fetal nucleated cells in maternal blood between the 18th and 22nd weeks of pregnancy using molecular cytogenetic techniques.

    Clinical genetics·2001
    Same author

    Captopril suppression versus salt loading in confirming primary aldosteronism.

    Hypertension (Dallas, Tex. : 1979)·2001

    Area of Science:

    • Clinical Chemistry
    • Analytical Instrumentation

    Background:

    • The Hitachi 705 is a new discrete analyzer designed for high-throughput clinical laboratory testing.
    • Evaluating new analytical instruments is crucial for ensuring reliable diagnostic capabilities.

    Purpose of the Study:

    • To assess the performance, reliability, and practicability of the Hitachi 705 discrete analyzer.
    • To evaluate the accuracy and precision of 11 common clinical chemistry assays performed on the instrument.

    Main Methods:

    • The Hitachi 705 analyzer was evaluated, focusing on its spectrophotometer, sample/reagent probes, and performance for 11 specific assays.
    • Precision (CV), linearity, and total analytical error were assessed for each assay.
    • Results were compared against medically allowable error limits.

    Main Results:

    • The instrument achieved a throughput of 180 tests per hour for up to 19 analytes.
    • Total imprecisions ranged from 0.14% to 8.0% across different components and assays.
    • Good linearity was observed for most tests, with acceptable analytical error, except for calcium and phosphorus.

    Conclusions:

    • The Hitachi 705 discrete analyzer demonstrates good reliability and practicability for a wide range of clinical chemistry assays.
    • While generally accurate, specific analytes like calcium and phosphorus may require further validation or method adjustment.