Stiffness of central venous catheters
Acta Anaesthesiologica Scandinavica
|April 1, 1983
Summary
Catheter stiffness significantly impacts thrombus formation after central venous cannulation. Lower stiffness catheters (below 16 X 10(-6) Nm2) show a reduced incidence of blood clot development.
Area of Science:
- Biomedical Engineering
- Materials Science
- Vascular Surgery
Background:
- Thrombogenesis is a significant complication of central venous cannulation.
- Catheter stiffness is hypothesized to be a key factor, but data are lacking.
Purpose of the Study:
- To develop a method for measuring central venous catheter stiffness.
- To quantify the stiffness of various catheter materials.
- To correlate catheter stiffness with thrombus formation.
Main Methods:
- A cantilever beam method (Grass Model DA-7) was used to measure catheter tip deflection.
- Stiffness (bending stiffness, EI) was averaged from six catheter sections.
- Twenty-seven central venous catheters (silicone, polyurethane, PVC, polyethylene, Teflon) were tested.
Main Results:
- Silicone, polyurethane, and PVC catheters had bending stiffness below 16 X 10(-6) Nm2.
- Polyethylene and Teflon catheters were significantly stiffer.
- Heparinization and radioopacity did not substantially affect stiffness.
- Lower stiffness catheters (< 16 X 10(-6) Nm2) demonstrated significantly less thrombus formation.
Conclusions:
- Catheter material and design influence bending stiffness.
- Catheter stiffness is a critical factor in central venous catheter-related thrombosis.
- Softer catheters (EI < 16 X 10(-6) Nm2) are associated with reduced thrombus formation.