Total lymphoid irradiation in the Wistar rat: technique and dosimetry
Abstract:
The technical and dosimetric aspects of total lymphoid irradiation (TLI) in the Wistar rat were evaluated as part of a set-up to develop a new model for tumor xenotransplantation. Information obtained from anatomical dissections, radionuclide imaging of the spleen, lymphography and chromolymphography was used to standardize the localization of lymph nodes, thymus and spleen. A practical advantage was found in the simultaneous irradiation through two portals cut out in a lead plate. The two portals encompassed the lymphoid tissue above and below the diaphragm. A specially designed masonite phantom was used to measure the dose distribution in the simulated target volumes. Ionization chamber dosimetry, thermoluminescence dosimetry and film densitometry were used for measuring exposure and absorbed dose. Irradiation was performed with 250 kV X rays (HVL 3.1 mm Cu). The dose rate was regulated by adjusting the treatment distance. The dose inhomogeneity measured in the target volumes varied between 80-100%. The side scatter dose to non target tissues under the shielded area between the two portals ranged between 20-30%. The technique and dosimetry of total lymphoid irradiation in Wistar rats are now standardized and validated and pave the way for tumor xenotransplantation experiments.