Collaborative evaluation of the microbial profile system for quantitative antimicrobial susceptibility testing.

E H Gerlach, R N Jones, A L Barry

    Journal of Clinical Microbiology
    |March 1, 1983
    PubMed
    Summary

    The Microbial Profile System (MPS) antimicrobial panels demonstrated high agreement with reference methods across multiple studies. This antimicrobial susceptibility testing system is reliable for clinical isolates and resistant strains.

    Related Concept Videos

    You might also read

    Related Articles

    Articles linked to this work by shared authors, journal, and citation graph.

    Sort by
    Same author

    Something New-Nitric Acid a Remedy for Stings and Poisons.

    The Southern medical record·2022
    Same author

    Cefuroxime, an in vitro Comparison with Six Other Cephalosporins.

    Proceedings of the Royal Society of Medicine·2010
    Same author

    Ceftazidime, a pseudomonas-active cephalosporin: in-vitro antimicrobial activity evaluation including recommendations for disc diffusion susceptibility tests.

    The Journal of antimicrobial chemotherapy·2009
    Same author

    Precision and accuracy of fluconazole susceptibility testing by broth microdilution, Etest, and disk diffusion methods.

    Antimicrobial agents and chemotherapy·2002
    Same author

    Standardization of broth microdilution and disk diffusion susceptibility tests for Actinobacillus pleuropneumoniae and Haemophilus somnus: quality control standards for ceftiofur, enrofloxacin, florfenicol, gentamicin, penicillin, tetracycline, tilmicosin, and trimethoprim-sulfamethoxazole.

    Journal of clinical microbiology·2001
    Same author

    Evaluation of daptomycin susceptibility testing by Etest and the effect of different batches of media.

    The Journal of antimicrobial chemotherapy·2001

    Area of Science:

    • Microbiology
    • Clinical Diagnostics
    • Antimicrobial Resistance

    Background:

    • Accurate antimicrobial susceptibility testing (AST) is crucial for guiding effective treatment of bacterial infections.
    • Microdilution methods are standard for determining minimum inhibitory concentrations (MICs).
    • Evaluating novel AST systems against established methods is essential for clinical adoption.

    Purpose of the Study:

    • To assess the performance and reliability of the Microbial Profile System (MPS) antimicrobial microdilution panels.
    • To compare MPS panel results with a reference microdilution method across various bacterial strains and study phases.

    Main Methods:

    • A three-center collaborative study involving three phases.
    • Phase I: Assessed inter- and intralaboratory agreement using strains with known susceptibility ranges.

    Related Experiment Videos

  • Phase II: Evaluated performance with strains exhibiting known resistance mechanisms and tested 359 clinical isolates.
  • Main Results:

    • High intralaboratory agreement (97.7% for MPS, 98.8% for reference) and interlaboratory agreement (96.2% for MPS, 96.0% for reference) were observed.
    • Overall agreement for strains with known resistance mechanisms was 94.7%, with specific high percentages for Enterobacteriaceae (94.5%), nonenteric gram-negative rods (95.4%), staphylococci (92.3%), and streptococci (96.6%).
    • Testing 359 clinical isolates yielded 97.3% agreement between MPS and the reference method.

    Conclusions:

    • The Microbial Profile System (MPS) antimicrobial microdilution panels demonstrated performance equivalent to the reference microdilution method.
    • MPS panels provide reliable and accurate results for a broad spectrum of bacteria, including resistant strains and clinical isolates.
    • The study supports the clinical utility of the MPS for antimicrobial susceptibility testing.