Multi-element analysis of the human body using neutron activation
Physics in Medicine and Biology
|March 1, 1983
Summary
This study calibrates a facility for in vivo neutron activation analysis to measure total body elements like nitrogen and potassium. The method accurately assesses elements in critically ill patients with minimal radiation exposure.
Area of Science:
- Nuclear medicine
- Medical physics
- Biomedical engineering
Background:
- Accurate measurement of total body elemental composition is crucial for assessing nutritional status and disease progression.
- In vivo neutron activation analysis (NAA) offers a non-invasive method for quantifying major elements in the human body.
Purpose of the Study:
- To describe the calibration of the Leeds in vivo neutron activation facility for simultaneous measurement of total body Na, Cl, P, Ca, N, and K.
- To incorporate the effects of body size and shape using a variable anthropomorphic phantom.
- To detail the correction methods for interfering nuclear reactions and determine the accuracy and precision of the technique.
Main Methods:
- Calibration of the Leeds in vivo NAA facility.
- Utilized a variable anthropomorphic phantom to account for body size and shape variations.
- Applied corrections for main interfering nuclear reactions.
- Determined accuracy and precision of the elemental measurements.
Main Results:
- The facility allows simultaneous measurement of total body Na, Cl, P, Ca, N, and K.
- Nitrogen (N) was measured with a coefficient of variation (CV) of 2.2% at 0.5 mSv dose equivalent.
- Potassium (K) was measured with a CV of 1.6% in the same counting period.
- Sodium (Na), Chlorine (Cl), and Phosphorus (P) were determined with CVs of 2.2%, 1.6%, and 2.9% respectively at 5 mSv.
- Calcium (Ca) was measured with a CV of 3% at 10 mSv dose equivalent.
Conclusions:
- The described calibration procedure is suitable for measuring elemental composition in critically ill patients.
- The technique provides accurate and precise measurements of total body elements with acceptable radiation doses.
- The method is advantageous as both irradiation and counting are performed with the patient in a supine position.