Related Experiment Video
Updated: Sep 21, 2026

Tubal Cytology of the Fallopian Tube as a Promising Tool for Ovarian Cancer Early Detection
Published on: July 25, 2017
Fletcher-Suit-Delclos gynecologic applicator: evaluation of a new instrument
Abstract:
A new instrument has been added to the Fletcher-Suit-Delclos group of gynecologic applicators. The colpostats can be used as mini-ovoids, but by adding a shielded cap, the instrument has the configuration of the original preload Fletcher colpostat. Dosimetric studies show that this applicator produced transmission ratios (the fraction of radiation transmitted through the tungsten shield) and isodose curves similar to the bladder trigone and anterior rectal wall of 10-25%. With the shield-containing cap removed, the mini-ovoid provides little reduction in dose to those areas and should be used with caution.