Related Experiment Video
Updated: Aug 25, 2026

Large Scale Non-targeted Metabolomic Profiling of Serum by Ultra Performance Liquid Chromatography-Mass Spectrometry (UPLC-MS)
Published on: March 14, 2013
High performance liquid chromatographic determination of caffeine in decaffeinated coffee, tea, and beverage products
Abstract:
A method was developed for determining caffeine in decaffeinated coffee, tea, and beverage products by high performance liquid chromatography (HPLC). The HPLC system consisted of a Bio-Sil ODS-5S C18 column, methanol-water (25 + 75) mobile phase at 1 mL/min, and a UV detector. The method is simple and specific. Caffeine recoveries were 93.8-98.3% and coefficients of variation were 0.90-2.25%.