Related Experiment Videos
Progressive dialysis encephalopathy from dialysate aluminum
Archives of Internal Medicine
|September 1, 1978
Summary
High aluminum levels in dialysis water caused progressive dialysis encephalopathy in patients. Deionizing the water significantly reduced aluminum, preventing further cases of this serious complication.
Area of Science:
- Nephrology
- Environmental Health Science
Background:
- Progressive dialysis encephalopathy is a serious complication of hemodialysis.
- High aluminum levels in water used for dialysis have been suspected as a cause.
Purpose of the Study:
- To investigate the link between water aluminum levels and dialysis encephalopathy.
- To assess the impact of water deionization on the incidence of this complication.
Main Methods:
- Observational study in a hemodialysis unit.
- Water quality analysis before and after deionization.
- Patient monitoring for neurological complications.
Main Results:
- Eight of 34 patients developed progressive dialysis encephalopathy.
- After deionization, water aluminum levels decreased from 0.637 mg/L to <0.001 mg/L.
- No further cases occurred in 29 patients post-deionization (P < .005).
Conclusions:
- Elevated aluminum in dialysis water is a significant risk factor for dialysis encephalopathy.
- Water deionization is an effective method to reduce aluminum levels and prevent this complication.