Heterophilic antibodies interfering with radioimmunoassay. A false-positive pregnancy test
Abstract:
A young man with amenorrhea had a consistently positive pregnancy test result (serum radioimmunoassay measurement of beta-human chorionic gonadotropin hormone). No fetal or placental tissue was found after uterine curettage and exploratory laparotomy. The false-positive pregnancy test result was due to heterophilic antibovine and antigoat antibodies in the patient's serum. These antibodies interfered with radioimmunoassays using goat antibodies. This case shows that serum heterophilic antibodies can interfere with immunoassays and result in unnecessary diagnostic procedures and/or unnecessary treatment.
More Related Videos
06:50A High-Throughput Electrochemiluminescence 7-Plex Assay Simultaneously Screening for Type 1 Diabetes and Multiple Autoimmune Diseases
Published on: May 29, 2020
03:33Rapid Detection of Fecal Antigen of Helicobacter pylori Infection Based on Double Antibody Sandwich Detection Technology
Published on: May 23, 2025