Presumed autoclave failure due to false-positive spore strip tests
Infection Control : IC
|September 1, 1982
Summary
Hospital autoclaves failed to kill spores due to a contaminated growth medium, Bacillus coagulans. A modified quality control incubation temperature is recommended to prevent future sterilization failures.
Area of Science:
- Microbiology
- Sterilization Processes
- Hospital Infection Control
Background:
- Autoclaves and ethylene oxide sterilizers are critical for hospital sterilization.
- Regular spore challenge testing verifies the efficacy of sterilization processes.
- Sterilization failures can lead to significant patient safety risks.
Purpose of the Study:
- To investigate the cause of apparent sterilization failures in hospital autoclaves.
- To identify factors contributing to false positive results in spore challenge tests.
- To recommend improvements in quality control procedures for sterilization monitoring.
Main Methods:
- Sterilization processes were challenged with Bacillus species spores.
- Autoclave performance was monitored using biological indicator test strips.
- Growth medium used for spore viability testing was analyzed for contaminants.
- Quality control procedures for growth medium incubation were reviewed.
Main Results:
- Seven autoclaves showed apparent failures on 18 of 46 test strips within a two-day period.
- No mechanical, physical, or human errors were identified as the cause of autoclave failures.
- Bacillus coagulans, a contaminant, was found in the growth medium, causing turbidity at 55°C.
- This contaminant interfered with the accurate assessment of spore kill in the tests.
Conclusions:
- Contamination of growth medium with Bacillus coagulans can lead to false positive sterilization test results.
- Standard quality control incubation at 37°C may not detect this specific contaminant.
- Implementing an additional quality control incubation at 55°C for trypticase soy broth (TSB) is recommended.
- This enhanced quality control measure can prevent future misinterpretation of sterilization efficacy.