Related Experiment Videos
In vitro analysis of self-shearing retentive pins
Abstract:
Combining stress, analysis, microscopic examination, mechanical testing of the shear mechanism, and retention of the Reten Pin leads to the following conclusions: 1. The suggested 0.006 inch pin-channel mismatch induces high lateral and apical stresses. Microscopically, this was seen to correlate with injury to the dentin. 2. The degree of retention was increased by using a smaller pin-channel mismatch. This correlates with smaller stresses and reduced dentinal damage. 3. The shear mechanism acts in a uniform manner, with a relatively small variation from the mean. It is suggested that for the best results the manufacturer should supply larger twist drills and pins with a somewhat deeper self-shearing groove to minimize apical involvement during shearing of the handle from the pin.