A solid sorbent personal sampling method for the determination of acrolein in air
Abstract:
A personal air sampling method, using activated carbon with subsequent analysis by gas chromatography, has been developed for acrolein in the 0.05 to 5 ppm (parts per million) range for a five liter air sample. The sensitivity is 0.25 microgram (microgram) of acrolein. This first known viable method for acrolein using a solid adsorbent was made possible by the discovery that acrolein can be adsorbed on, and recovered from, hydroquinone-treated carbon with efficiencies as high as 90 percent. Samples must be analyzed the same day or frozen and analyzed within five days. Airborne contaminants possibly present in the acrolein production area do not interfere with the analysis.
More Related Videos
08:33Use of Capillary Aerosol Generator in Continuous Production of Controlled Aerosol for Non-Clinical Studies
Published on: April 12, 2022
09:19Capturing Actively Produced Microbial Volatile Organic Compounds from Human-Associated Samples with Vacuum-Assisted Sorbent Extraction
Published on: June 1, 2022
Related Concept Videos
Sampling Methods: Sample Types
Solid samples include a variety of substances, such as sediments from water bodies, soil, metals, and biological tissues. Two standard methods for extracting sediments from water bodies are grab sampling and piston coring. Grab sampling involves using a device to collect a discrete sediment sample from the bottom of a water body with minimal disturbance. Grab samples do not always represent the entire area due to...
Sample Preparation for Analysis: Overview
Bulk or large solid samples are typically reduced in size using grinding, crushing, or milling techniques to increase the...
Atomic Absorption Spectroscopy: Atomization Methods
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing nebulizer...