Rapid amylase and lipase determinations by nephelometry
American Journal of Clinical Pathology
|July 1, 1978
Summary
The Coleman 91 nephelometer offers fast amylase and lipase testing for emergencies. While generally accurate, extremely lipemic samples require dilution for reliable results.
Area of Science:
- Clinical Chemistry
- Biochemical Assays
Background:
- Emergency laboratory testing requires rapid and reliable diagnostic tools.
- Amylase and lipase assays are crucial for diagnosing pancreatic conditions.
Purpose of the Study:
- To evaluate the performance of the Coleman 91 nephelometer for amylase and lipase assays.
- To assess its suitability for emergency diagnostic requests.
Main Methods:
- The Coleman 91 nephelometer was used for amylase and lipase assays.
- Assay linearity, precision, and normal ranges were compared to established methods (Phadebas, amyloclastic, titrimetric).
- The impact of lipemic sera on assay results was investigated.
Main Results:
- Amylase assay showed favorable linearity and precision comparable to Phadebas assay in serum and urine.
- Serum amylase normal ranges were slightly higher than amyloclastic methods.
- Lipase assay demonstrated improved linearity over titrimetric methods but variable kinetics.
- Lipase assay precision was lower than amylase, with higher normal ranges than titrimetric methods.
- Falsely low amylase and lipase results were observed in highly lipemic sera, necessitating dilution.
Conclusions:
- The Coleman 91 nephelometer is a viable option for rapid amylase and lipase testing, especially in emergency settings.
- Careful interpretation is needed for lipemic samples, requiring serial dilution for accurate quantification.