Modified Jendrassik--Grof method for bilirubins adapted to the Abbott Bichromatic Analyzer
Clinical Chemistry
|October 1, 1978
Summary
A new semi-micro method accurately measures total and direct bilirubins using only 75 microliters of sample. This simple, rapid assay offers comparable results to existing methods, with good precision and linearity.
Area of Science:
- Clinical Chemistry
- Analytical Biochemistry
Background:
- Bilirubin levels are critical indicators of liver function and hemolytic disease.
- Accurate and efficient bilirubin determination is essential for timely clinical diagnosis.
- Existing automated methods may require larger sample volumes or lack specific advantages.
Purpose of the Study:
- To develop and validate a semi-micro method for quantifying total and direct bilirubins.
- To assess the method's performance characteristics, including accuracy, precision, and linearity.
- To address potential interferences in bilirubin assays.
Main Methods:
- Utilized a discrete analyzer (Abbott ABA-100) with a 75 microliter sample volume.
- Employed a serum blank and measured azobilirubin color development at 550 nm.
- Investigated and mitigated interference from hemoglobin and its derivatives using caffeine and acid blanks.
Main Results:
- Achieved comparable results to the automated Jendrassik and Grof procedure (total r = 0.999, direct r = 0.990).
- Demonstrated linearity up to 300 mg/liter.
- Showcased good day-to-day precision (CV 8.8% and 4.7% for different concentrations).
Conclusions:
- The described semi-micro method provides a simple, rapid, and accurate means for determining total and direct bilirubins.
- The method is suitable for clinical laboratory use, offering a low sample volume requirement.
- Effective strategies were implemented to overcome common assay interferences.