Quantitative electron probe x-ray microanalysis of biological bulk specimens
Summary
This tutorial details quantitative X-ray microanalysis for thick biological specimens. Organic standards and relative intensities (P/B-ratios) offer advantages for accurate elemental analysis in bulk samples.
Area of Science:
- Analytical Chemistry
- Materials Science
- Biophysics
Background:
- Quantitative X-ray microanalysis of thick biological specimens presents challenges compared to thin sections.
- Existing quantitative methods require further development for bulk biological samples.
Purpose of the Study:
- To provide a detailed examination of ZAF-corrections for bulk biological specimens.
- To address challenges in quantitative elemental analysis using X-ray microanalysis.
- To compare different types of standards and intensity measurement methods.
Main Methods:
- Detailed examination of ZAF-correction components and the ratio model.
- Investigation of secondary fluorescence effects from the continuum in metal analysis (Z>20).
- Comparison of organic (gelatin/resin) and crystal standards for homogeneity and quantitation.
Main Results:
- Organic standards demonstrate superior homogeneity and quantitation performance compared to crystal standards.
- Secondary fluorescence by the continuum is a significant factor in metal analysis of biological bulk specimens.
- Relative intensities (P/B-ratios) generally offer advantages over absolute X-ray intensities for quantitative analysis.
Conclusions:
- ZAF-correction methods and standard selection are critical for accurate quantitative X-ray microanalysis of biological bulk specimens.
- Organic standards and P/B-ratios are recommended for improved quantitation in this field.
- Further research may be needed for specific cases where P/B-ratios are not advantageous.