Related Experiment Video
Updated: Aug 11, 2026

Safety Precautions and Operating Procedures in an (A)BSL-4 Laboratory: 2. General Practices
Published on: October 3, 2016
Problems in the disinfection of class 1 microbiology safety cabinets
Abstract:
Microbiology safety cabinet disinfection procedures using formaldehyde have been tested. Tubercle bacilli were killed by concentrations of formaldehyde obtained by heating commercial formalin irrespective of whether the bacilli were in the cabinet free space or above the prefilters. However, Bacillus stearothermophilus spore papers for for the testing of low temperature steam/formaldehyde sterilisers were almost never sterilised and a strain of Staphylococcus epidermidis (NCTC 7944) showed a resistance intermediate between the B stearothermophilus spores and the tubercle bacilli. Tests using a vaccine strain of poliovirus type 3 indicated a considerable degree of resistance of the virus to the action of formaldehyde. No such resistance was demonstrated by vaccinia virus or echovirus 14. Chemical and biological evidence is presented which indicates that filter paper discs are an unsuitable carrier material for a challenge organism in testing the efficiency of any formaldehyde sterilising process. Recommendations are made towards developing a satisfactory test procedure.
Related Concept Videos
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Physical Methods for Controlling Microbial Growth: Radiation and Filtration
Chemical Agents for Microbial Control
Key Techniques in Microbiology