Comparison of two methods for same-day identification of Enterobacteriaceae
American Journal of Clinical Pathology
|September 1, 1982
Summary
Two rapid commercial methods, API 20E and Micro-ID, offer same-day enterobacterial identification. Micro-ID showed higher accuracy than API 20E for clinical isolates, providing valuable rapid results.
Area of Science:
- Clinical microbiology
- Bacterial identification
- Diagnostic methods
Background:
- Accurate and rapid identification of Enterobacteriaceae is crucial for effective patient treatment.
- Traditional overnight methods can delay critical clinical decisions.
- Need for validated same-day identification systems in clinical laboratories.
Purpose of the Study:
- To evaluate the performance of two commercial same-day identification methods, API 20E and Micro-ID.
- To compare their accuracy and speed in identifying clinically isolated Enterobacteriaceae.
- To assess the clinical utility of rapid identification results.
Main Methods:
- Evaluation of 368 clinical isolates of Enterobacteriaceae using API 20E and Micro-ID systems.
- Analysis of identification accuracy at species and genus levels.
- Comparison of results at specific time points (API at 5 hours, Micro-ID at 4 hours).
Main Results:
- Micro-ID achieved 90.0% correct species identification at 4 hours, superior to API 20E's 78.5% at 5 hours.
- Both methods showed excellent identification of Escherichia coli.
- API 20E struggled with Serratia, Citrobacter, and Providencia, while Micro-ID had minor inaccuracies with Serratia marcescens and Klebsiella pneumoniae.
Conclusions:
- Both API 20E and Micro-ID offer rapid, albeit less accurate, identification compared to overnight methods.
- Micro-ID demonstrates higher overall accuracy for same-day enterobacterial identification.
- These rapid systems provide valuable clinical information, balancing speed with a slight sacrifice in accuracy.