An evaluation of a plastic strip thermometer
JAMA
|January 15, 1982
Summary
The Clinitemp fever detector is an unreliable tool for detecting fevers. This plastic strip thermometer failed to accurately identify fevers when compared to traditional mercury/glass thermometers.
Area of Science:
- Medical Devices
- Clinical Thermometry
- Diagnostic Accuracy
Background:
- Forehead thermometers offer a non-invasive method for temperature assessment.
- Plastic strip thermometers, like the Clinitemp, aim to provide a convenient fever detection tool.
- Evaluating the accuracy of alternative thermometers against established standards is crucial for patient care.
Purpose of the Study:
- To assess the diagnostic accuracy of the Clinitemp fever detector.
- To compare the Clinitemp's fever detection capabilities against mercury/glass thermometers.
- To determine if the Clinitemp is a suitable substitute for traditional thermometers.
Main Methods:
- A study involving 613 patients compared Clinitemp readings to mercury/glass thermometer measurements.
- Different temperature thresholds (37.8°C/100°F and 37.2°C/99°F) were evaluated for fever detection using the Clinitemp.
- The impact of ambient air temperature on Clinitemp accuracy was considered.
Main Results:
- At a threshold of 37.8°C (100°F), the Clinitemp detected only 34% of fevers identified by mercury/glass thermometers.
- Sensitivity increased to 0.81 when a lower threshold of 37.2°C (99°F) was used.
- Ambient temperature variations significantly affected the accuracy of the Clinitemp readings.
Conclusions:
- The Clinitemp fever detector demonstrates poor accuracy in identifying clinically significant fevers.
- Its performance is inconsistent and influenced by environmental factors.
- The Clinitemp is not a reliable substitute for traditional mercury/glass thermometers in fever detection.