Methods for routine calibration of brachytherapy sources

J F Williamson, F M Khan, S C Sharma

    Radiology
    |February 1, 1982
    PubMed
    Summary

    Routine calibrators for brachytherapy sources show significant deviations. Calibration accuracy is affected by geometry and capsule filtration, necessitating a new mathematical correction and revised assay procedures for reliable dose measurements.

    Related Concept Videos

    You might also read

    Related Articles

    Articles linked to this work by shared authors, journal, and citation graph.

    Sort by
    Same author

    Establishing CBT-T in an NHS moderate-severe adult eating disorder service.

    Eating disorders·2025
    Same author

    Quantitative verification of <sup>192</sup> Ir PDR and HDR source structure by pin-hole autoradiography.

    Medical physics·2018
    Same author

    HLA-DQB1*05:144, a novel allele, discovered in a Southeast Asian donor for stem cell transplantation.

    HLA·2017
    Same author

    WE-E-213CD-05: A Non-Rigid Image Registration Algorithm That Accommodates Organ Segmentation Error.

    Medical physics·2017
    Same author

    Factors responsible for increased percent recirculation in arterio-venous fistula among the haemodialysis patients.

    Bangladesh Medical Research Council bulletin·2013
    Same author

    On the biological basis for competing macroscopic dose descriptors for kilovoltage dosimetry: cellular dosimetry for brachytherapy and diagnostic radiology.

    Physics in medicine and biology·2013

    Area of Science:

    • Medical Physics
    • Radiation Oncology
    • Nuclear Medicine

    Background:

    • Accurate calibration of brachytherapy sources is crucial for effective radiation therapy.
    • Routine calibrators may introduce errors due to geometric and filtration effects.
    • Existing calibration procedures may not be suitable for brachytherapy sources.

    Purpose of the Study:

    • To compare the accuracy of two routine geometry calibrators against open-air measurements for brachytherapy sources.
    • To identify and quantify the sources of error in dose calibration.
    • To develop a method for correcting geometric and filtration effects.

    Main Methods:

    • Comparison of a nuclear medicine dose calibrator and a custom jig with open-air measurements.
    • Evaluation of 137Cs, 192Ir, and 226Ra brachytherapy sources.
    • Analysis of geometric effects, buildup, and inverse square law deviations.

    Main Results:

    • The custom jig showed significant deviation (5%-15%) from the inverse square law due to scattering.
    • Dose calibrator response varied with source energy and capsule thickness.
    • Filtration differences accounted for 27% of calibration factor variation.

    Conclusions:

    • A mathematical logarithm was developed to correct for geometric and filtration effects.
    • Manufacturer-provided activity assay procedures are inadequate for brachytherapy sources.
    • Revised calibration and assay methods are necessary for accurate brachytherapy dosimetry.